Biocompatibility is a critical requirement in the selection of materials for biomedical implants. Among the most extensively studied materials, Nitinol (NiTi)—an equiatomic nickel–titanium alloy—has surpassed conventional implant materials such as stainless-steel owing to its unique properties, including the shape memory effect, super elasticity, and excellent corrosion resistance [1,2]. In biomedical applications, NiTi is widely used in devices such as vascular stents, orthopedic implants, and guidewires [3].

The transition from conventional fabrication routes, such as casting and sintering, to additive manufacturing (AM) offers significant advantages, including the production of complex geometries and reduced material waste. However, additive manufacturing of NiTi remains challenging due to rapid solidification rates and the potential formation of brittle secondary intermetallic phases (e.g., TiNi₃ and Ti₂Ni), which can adversely affect mechanical performance and corrosion resistance.

This study investigates the effect of thermal treatment—specifically base-plate preheating—on the corrosion behavior of in-situ alloyed and pre-mixed NiTi fabricated by Laser Direct Metal Deposition (LDMD), with comparisons made to conventionally cast NiTi. Corrosion performance was evaluated using potentiodynamic polarization testing in Hank’s solution to simulate physiological conditions, in accordance with ASTM G59. The results indicate that non-preheated, in-situ alloyed NiTi exhibits superior corrosion resistance compared to preheated samples, pre-mixed NiTi, and as-cast counterparts

Contributing Authors

  • Samuel S Skhosane
    CSIR
  • Nthabiseng N Maledi
    University of the Witwatersrand
  • Sisa S Pityana
    CSIR
  • Michael M Bodunrin
    University of the Witwatersrand
  • Samuel S Oke
    University of the Witwatersrand

Presenter

Samuel S Skhosane

CSIR

Samuel S Skhosane

CSIR

About Samuel

I am a welding engineer at the Council for Scientific and Industrial Research (CSIR) in South Africa. I hold an MSc in Welding Metallurgy from the University of the Witwatersrand and am currently pursuing a PhD. My MSc dissertation focused on investigating the effect of niobium (Nb) on the mechanical properties of laser coated Nitinol (NiTi) for the surface modification of Ti6Al4V.

To date, I have published 22 research articles and contributed two book chapters. My expertise lies in laser additive manufacturing and processing of materials using laser. I am passionate about my career to drive innovation in materials engineering.

Track: Laser Additive Manufacturing
Topic:
Biomedical
Session: Process and Materials Performance in metal AM
Date/Time:
Location:

Keywords

  • Biocompatibility
  • Biomedical
  • Corrosion
  • Nitinol
  • Pre-Heat